Rule 44:67:05 CONSTRUCTION STANDARDS

CHAPTER 44:67:05

 

CONSTRUCTION STANDARDS

Section

44:67:05:01        Submittal of plans.

44:67:05:02        Procedure rooms and services.

44:67:05:03        Recovery rooms and services.




Rule 44:67:05:01 Submittal of plans.

          44:67:05:01.  Submittal of plans. No construction may begin until plans and specifications covering the construction of new buildings and additions or material alterations to existing buildings are approved by the department. A written narrative describing the intended use of the proposed construction must accompany the plans and specifications.

 

          Source: 33 SDR 107, effective December 26, 2006.

          General Authority: SDCL 34-23A-51.

          Law Implemented: SDCL 34-23A-51(1),(3),(5),(9).

 




Rule 44:67:05:02 Procedure rooms and services.

          44:67:05:02.  Procedure rooms and services. Each abortion facility shall be designed and equipped for the types of abortion procedures to be performed.

 

          Procedure rooms shall meet the following requirements:

 

          (1)  A minimum area of 115 square feet (10.8 square meters);

          (2)  Equipped with a vinyl floor surface;

          (3)  A work station and handwashing sink. Handwashing facilities must be equipped with hands-free controls and have a water supply spout mounted so that the discharge is a minimum five inches above the rim of the fixture;

          (4)  Adequate lighting with protective covers to prevent lamps from shattering; and

          (5)  Battery operated emergency lighting.

 

          Procedure services shall include:

 

          (1)  Provisions for immediate post-procedure care; and

          (2)  Emergency drugs and supplies specified by the medical staff.

 

          Source: 33 SDR 107, effective December 26, 2006.

          General Authority: SDCL 34-23A-51.

          Law Implemented: SDCL 34-23A-51(5).

 




Rule 44:67:05:03 Recovery rooms and services.

          44:67:05:03.  Recovery rooms and services. Each abortion facility shall be designed and equipped for the types of abortion procedures to be performed.

 

          Recovery rooms shall meet the following requirements:

 

          (1)  A minimum area of 45 square feet in each recovery station;

          (2)  Privacy curtains surrounding each recovery station;

          (3)  A work station and handwashing sink. Handwashing facilities shall be equipped with hand-free controls and have a water supply spout mounted so that the discharge is a minimum of five inches above the rim of the fixture;

          (4)  Adequate lighting with protective covers to prevent lamps from shattering;

          (5)  An emergency call system to summon staff; and

          (6)  Battery operated emergency lighting.

 

          Recovery services shall include:

 

          (1)  Provisions for immediate post-procedure care; and

          (2)  Emergency drugs and supplies specified by the medical staff.

 

          Source: 33 SDR 107, effective December 26, 2006.

          General Authority: SDCL 34-23A-51.

          Law Implemented: SDCL 34-23A-51(5).

 